BOD, COD and Solids in Wastewater: APHA and IS Method Framework

Understand BOD, COD and total suspended solids testing using APHA Standard Methods and current IS 3025 references, sampling, preservation and trend interpretation.

All technical articlesDiscuss a testing requirement
Technical illustration for BOD, COD and Solids in Wastewater: APHA and IS Method FrameworkChem-Tech Laboratories

Article overview

What this technical guide covers

Use the section guide to move directly to the operating problem, evidence or testing decision most relevant to your requirement.

  1. 02What the programme must answer
  2. 03BOD, COD and TSS each respond to different system changes
  3. 04Standards and method framework
  4. 05Sampling and pre-analytical control

Technical perspective

BOD, COD and suspended solids are foundational wastewater indicators, but they describe different properties. Biochemical oxygen demand estimates oxygen consumed by biologically degradable material under specified test conditions; chemical oxygen demand measures oxidizable material under the selected chemical digestion method; and total suspended solids quantify non-filterable residue under a defined gravimetric procedure. They should not be used interchangeably.

For Indian wastewater programmes, the IS 3025 series provides current test methods, including IS 3025 (Part 17):2022 for non-filterable residue or total suspended solids and IS 3025 (Part 44):2023 for BOD (Second Revision). Current BIS listings should be checked for COD under IS 3025 (Part 58) and other parameters. APHA/AWWA/WEF Standard Methods, 24th edition provides a widely used international method framework for water and wastewater.

02

Article section

What the programme must answer

Testing or auditing is valuable only when the result is tied to a defined decision. The questions below should be agreed before sampling so that the sample, method, specification and response are aligned.

01

Treatment loading

Use COD, BOD and solids to characterize influent strength and loading variability.

02

Process performance

Compare influent, intermediate and final effluent to understand removal efficiency and biological-treatment behaviour.

03

Compliance monitoring

Test final discharge against the permit, consent or contract specification using the required method and sampling basis.

04

Troubleshooting

Use changing BOD/COD relationships, solids and process indicators to investigate toxicity, bypass, biomass loss or source changes.

03

Article section

BOD, COD and TSS each respond to different system changes

COD is generally faster than BOD and can be a useful operational indicator, but it is not a direct substitute unless a site-specific relationship has been established. BOD depends on biological activity and can be affected by toxicants, dilution, seed conditions and nitrification controls under the specified method. TSS can rise from poor clarification, hydraulic overload, sludge carryover or particulate influent changes.

Ratio trends can be more informative than one number. A changing BOD-to-COD relationship can suggest a shift in biodegradability or influent composition. Rising effluent TSS with stable dissolved-load indicators can point toward solids separation rather than biological oxidation. Laboratory data should be paired with flow, aeration, sludge, chemical-dosing and operational logs.

  • Define whether samples are grab or time/flow composites according to monitoring objective and permit requirement.
  • Measure or record flow with concentration when mass loading is important; concentration alone can hide higher total discharge at increased flow.
  • Record pH, temperature and unusual influent events that can explain BOD or process behaviour.
  • Use consistent sampling points before and after treatment stages for performance calculations.
  • Do not infer regulatory compliance from internal operational samples if mandated sampling basis differs.
04

Article section

Standards and method framework

Current IS and APHA methods should be selected according to the required reporting framework. The exact method conditions, sample preservation, interference treatment and quality controls must be followed; this article identifies the framework rather than reproducing copyrighted procedures. APHA's 24th edition is the current print edition and APHA directs users to its online Standard Methods service for the most up-to-date methods.

Reference / methodRole in the testing planImportant use condition
IS 3025 (Part 17):2022Total suspended solids / non-filterable residueCurrent gravimetric suspended-solids method listed by BIS.
IS 3025 (Part 44):2023Biochemical Oxygen DemandCurrent second-revision BOD method listed by BIS.
IS 3025 (Part 58) – verify current BIS revisionChemical Oxygen DemandUse current revision required by specification or accreditation.
APHA Standard Methods, 24th ed.BOD, COD, solids and extensive water/wastewater parametersSelect specific Standard Method accepted by regulator/customer.
IS 17614 / ISO 5667 sampling-preservation familyWater sampling and preservation guidanceApply exact part relevant to sample and objective.
APHA/Standard Methods 5210 B (24th ed./current online)Five-day biochemical oxygen demand (BOD) frameworkUse exact current method conditions, dilution/seed controls and the method required by the consent or customer.
APHA/Standard Methods 5220 D (24th ed./current online)Closed reflux, colorimetric COD methodUse where accepted; control chloride/interferences, digestion, calibration and method-specific QC.
APHA/Standard Methods 2540 D (24th ed./current online)Total suspended solids dried at 103–105 °CUse the current method and representative solids sampling/subsampling.

A method number is not, by itself, a product specification or legal limit. The method tells the laboratory how a defined measurand is determined; the applicable regulation, product standard, consent, customer specification or validated internal plan supplies the criterion and decision context.

05

Article section

Sampling and pre-analytical control

Wastewater can change rapidly between collection and analysis. Biological activity continues in the bottle, solids settle, volatile components can be lost and some analytes transform. Sampling and preservation therefore need as much control as laboratory analysis. The sampling plan should capture production cycles and hydraulic variation rather than a single convenient moment when the plant is performing well.

Composite samples need mixing and subsampling procedures that preserve solids representativeness. Grab samples can be appropriate for some events or permit conditions, but they should not be described as a daily average when the discharge varies strongly through the production cycle.

  • Use the required grab or composite approach and document start/end times and flow basis where relevant.
  • Mix composite containers appropriately while preventing loss of representative solids before subsampling.
  • Cool or preserve samples according to the specified method and analyse within applicable holding time.
  • Record production shutdowns, CIP or cleaning, chemical dumps or storm inflow that affect representativeness.
  • Use dedicated containers for parameters with incompatible preservatives.
06

Article section

How to interpret results and trends

Trend interpretation should combine concentration, removal efficiency and load. A treatment plant can show stable effluent concentration while influent strength changes dramatically; removal percentage can look excellent when influent is very high but final discharge still fails; and low flow can make a concentration spike less important to total load than it appears. The governing consent or contract determines the actual compliance metric.

Operational correlations should be established from local data, not copied from another plant. If COD is used as a rapid surrogate for BOD, the site should demonstrate and periodically review the relationship for its own wastewater because process changes can alter biodegradability.

Observed result / patternTechnical interpretationManagement response
COD rises in influent while BOD/COD ratio fallsMore refractory load or toxicity is possibleReview production chemicals, cleaning streams and raw-material changes.
Effluent TSS rises while COD/BOD changes lessClarifier or solids-separation issueReview sludge blanket, hydraulics, flocculation and return sludge.
BOD unexpectedly low relative to CODToxic inhibition or low biodegradability possibleReview method QC, toxicity and influent chemistry.
All effluent parameters rise after high flowHydraulic overload or bypass possibleReview equalization, storm or CIP events.
Stable concentration but discharge volume risesMass load can increaseCalculate load and compare with consent conditions.
07

Article section

Build the result into a control programme

A wastewater monitoring programme should include control charts for influent and effluent, operational trigger levels and event annotations. It should distinguish regulatory samples from process-control samples and ensure that methods used for formal reporting match consent or customer requirements.

Trend BOD, COD, TSS and flow together. When results change, staged sampling through the treatment train can locate where performance is lost more effectively than repeatedly testing only the final outlet.

  • Trend BOD, COD, TSS and flow together rather than separately.
  • Annotate shutdowns, CIP, chemical dosing and major production changes on trend charts.
  • Use staged sampling after an adverse result to locate the treatment step losing control.
  • Verify method and edition required by pollution-control consent or contract before formal reporting.
  • Rebaseline treatment performance after capacity expansion or major process change.
08

Article section

References and related Chem-Tech resources

Standards, regulations and official methods are revised over time. Before a result is used for a regulatory conformity decision, confirm the current FSSAI notification or Gazette text, the exact food or water category, the latest applicable BIS edition and amendments, the customer or statutory specification, the sampling requirement and the laboratory method/scope. FSSAI compendia are useful ready references, but the original notifications and amendments prevail where there is a conflict. This article is technical guidance, not a substitute for the applicable law, permit or product specification.